SISTEM: 1.5T Excite II (Versi Piranti Lunak. 11.0M4)
1. Sistem iki 10.0M4, sawise upgrade kanggo Excite II 11.0M4, 8 saluran NV lan awak ora bisa mindai
2.Pesen kesalahan: Koil RF sing diwènèhaké ora cocog karo coil sing dipilih.
1. Copot 8ch NV lan awak coil saka 'situs coil', miwiti maneh piranti lunak, banjur nambah 8ch NV lan awak saka 'kasedhiya coil' kanggo 'situs coil', restart software, 8ch NV OK, nanging masalah padha dumadi karo 8ch awak.
2.Test awak 8ch ing situs liyane (Excite II), scan OK.
3.LFC kanggo Shanghai PLA 455, isih ora bisa
4.Mriksa log kesalahan: Koil RF sing diwènèhaké ora cocog. Status sambungan kumparan ndhuwur: 0, Kode kumparan ndhuwur: Null, Status sambungan kumparan ngisor: 1, Kode kumparan ngisor: 7t8jkd98cg58919otdqaft7c0bwilvc8, jeneng kumparan Rx: 8Ch Badan Ngisor, Kode kumparan Rx: 7t8jkd98cgwibqaft
5.Sunting 8ch Ndhuwur, Ngisor, lan kode koil FOV Lengkap '7t8jkd98cg58919otdqaft7c0bwi1vc8' minangka '7t8jkd98cg58919otdqaft7c0bwilvc8'. Wiwiti maneh piranti lunak, pindai OK!
Amarga 8ch NV lan awak coil ora didhukung ing 10.0M4. Konfigurasi loro kumparan iki diowahi kanthi manual ing Excite I, konfigurasi ora duwe "kode kumparan". Sawise nganyarke, konfigurasi coil sing dipulihake bisa uga ora padha karo 11.0 M4.